产品概述
Ultra low range turbidity measurement.
Filtertrak Laser Nephelometer offers the best resolution in the market. Detects changes in turbidity as low as 0.0005 NTU. Complies with USEPA approved Hach Method 10133 for regulatory reporting and compliance monitoring. Compatible with Hach Multi-Sensor, Multi-Parameter Digital Controllers, sensor can be used with Hach's new SC200 or SC1000 Digital Controllers. Includes Manual and Stablcal Turbidity Standards.
The FT660 is passing the torch to the TU5400 sc. The last day to order the FT660 is July 27, 2018. See the recommended products section of this page or visit HachTU5.com to learn more.
主要特点
- Lowest Limit of Detection in the Market:Using advanced laser optics and signal processing, the instrument detects increased concentrations of submicron-sized particles that are a precursor to larger particles. This allows for early filter deterioration detection that meets or exceeds that of particle counters all with the day-to-day convenience, simplicity, and reliability of a Hach turbidimeter. Operators can detect impending filter breakthrough, delineate filter ripening, and maximize effective filter run time. The Hach Filtertrak 660 sc Laser Nephelometer has the ability to detect a 0.3 mNTU or 0.0003 NTU change in turbidity, providing operators with confidence in their turbidity measurement.
- Meets All USEPA Filter Monitoring Requirements:The Filtertrak sensor uses U.S. Environmental Protection Agency (USEPA) approved Hach Filtertrak Method #10133, the method specified in the USEPA LT2 draft guidance manual for compliance monitoring of membrane filters. The sensor meets all regulatory requirements for individual filter and combined effluent monitoring.
- Monitor, Optimize, and Report:USEPA method compliance means that treatment plants with conventional or membrane filtration can monitor and optimize, as well as report, using this single technology.
- Compatible with Hach Multi-Sensor, Multi-Parameter Digital Controllers:The Filtertrak sensor can be used with any of Hach's sc Digital Controllers. Each sc controller accepts from two to eight sensors. Multiple controllers can be networked to accommodate many more sensors and parameters, reducing the cost per measuring point. Just plug in any Hach “plug and play digital sensor and it's ready to use without software configuration. “Plug and play connectivity means there's no complicated wiring or set up. The Filtertrak 660 sc Laser Nephelometer controller system now can manage multiple digital sensors.
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | (Defined according to ISO 15839) From ±3% of reading or ±5 mNTU, whichever is greater |
| Alarm | Three set-point alarms Two nephelometric set-point alarms (SOM required) |
| Cable Length | 1.8m (5.9 feet) |
| Calibration Method | Stablcal, stabilized formazin |
| Certifications | CE certified by Hach Company to EN61326 (industrial levels) |
| Certifications | Listed by ETL to UL 61010A-1: Certified by ETL to CSA C22.2 No. 1010.1: CE certified by Hach Company to EN 61010-1 |
| Communication | MODBUS® RS485/ RS232, PROFIBUS DP, HART available with sc controller |
| Controller | Dual Digital |
| Dimensions (H x W x D) | 406 mm x 305 mm x 254 mm |
| Display Resolution | 0.001 mNTU up to 9.9999 mNTU |
| Fitting Type | Sample Inlet: 1/4-inch NPT female, 1/2-inch hose barb (provided); Drain: 1/2-inch NPT female, 1/2-inch hose barb (provided), clear drain tubing is recommended |
| Flow | 100 - 750 mL/min |
| Flow Rate | 100 - 750 mL / min (1.6 to 11.9 gal/hour) |
| Light Source | 660 nm Laser Diode |
| Lower Limit of Detection (LOD) | (Defined according to ISO 15839) Less than 0.3 mNTU |
| Lowest Expected Reading | 5 mNTU using deionized, reverse osmosis water, based on statistical averages from three instruments |
| Measurement Frequency | One reading per second |
| Measuring Principle | Nephelometric light scatter at 90 degrees relative to incident monochromatic light beam |
| Mounting | Wall and floor stand |
| Number of Channels | 2 |
| Operating Humidity | 5 - 99 % non-condensing |
| Operating Temperature Range | 0 - 40 °C |
| Power Requirements (Voltage) | 24 VDC |
| Range | (0.0 - 5.0 NTU) 5000 mNTU = 5.000 NTU |
| Repeatability | + |
| Repeatability | (Defined according to ISO 15839) Better than ± 3.6 % of reading at 24 mNTU; as RSD (or as coefficient of variation) |
| Resolution | 0.01 mNTU from 10.00 - 99.99 mNTU |
| Resolution | 0.1 mNTU from 100.0 - 999.9 mNTU |
| Resolution | 1 mNTU from 1000 - 5000 mNTU |
| Sample Input | 0.25 " NPT female thread (1/4" compression fitting provided) |
| Sample Output | 0.5 " NPT female thread (1/2" hose barb provided) clear drain tubing is recommended |
| Sample Temperature | 0 - 50°C (32 - 122°F) |
| Signal Average Time | User selectable from 0, 6, 30, 60, 90 seconds; default 90 seconds |
| Storage Conditions | -20 °C to 60 °C |
| Weight | 4.54 kg |
| What's included? | Hach sc200 Controller, Mounting Hardware, sc200 Basic User Manual, FT660sc, FT660sc User Manual, and 1000 mL 800 mNTU Stablcal ®. (The controller power cord is ordered separately.) |
资料下载
- Monitoring Iron Transport in Power Generation- Application Note
- The Use of Simplified Statistical Processing Techniques to Increase Sensitivity of Detection to Particulate Breakthrough
- Monitoring Iron Transport in the Steam Cycle via Grab Sample and Online Methods
- Further Advances in Monitoring Low Level Iron in the Steam Cycle
- Monitoring Iron Transport in the Steam Cycle via Grab Sample and Online Methods
- Laser Nephelometer Calibration Guide for Monitoring Flow Accelerated Corrosion and Treatment Efficiency in the Steam Cycle
- Monitorización del transporte de hierro en el ciclo de vapor mediante muestras en el ciclo de vapor mediante muestras
- Introduction to Laser Nephelometry: An Alternative to Conventional Particulate Analysis Methods, by Mike Sadar, Hach Company
Case Studies
Municipal Wastewater Treatment · Controllers and Sensors — Application Case
Site Conditions
This is a Municipal Wastewater Treatment customer whose plant is at the inlet, biological and outlet stages of a municipal plant. The main task was online water monitoring with a controller and sensor combination.
The Problem
The difficulty lies in high solids and corrosive fractions, with large diurnal flow swings; at the same time effluent parameters drift and probes foul with sludge and biofilm. Specification therefore means matching each controller channel to its sensor type.
What We Did
We took a Hach channel approach — supplying against the Hach original part numbers. Delivery was built around matching each controller channel to its sensor type, with point-by-point checks on: stable compliance with Grade 1A discharge limits and online data reporting.
Operating Outcome
In service the system performs as intended — factory-matched combinations that make faults easier to locate, with routine maintenance falling noticeably.
Food & Beverage · Controllers and Sensors — Application Case
Application Context
dosing, pasteurising, filling and CIP stages is a typical Food & Beverage duty. The project started from Controllers and Sensors, focusing on online water monitoring with a controller and sensor combination.
Process Constraints
frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles, and syrups, dairy and beverages that scale easily, under strict hygienic requirements only amplify it. The Controllers and Sensors specification must therefore cover matching each controller channel to its sensor type.
Delivery Approach
Delivery was handled as the Hach channel — supplying against the Hach original part numbers. Execution centred on matching each controller channel to its sensor type, together with food-safety and hygienic design per 3-A / EHEDG.
Field Results
Operation shows the benefit clearly — factory-matched combinations that make faults easier to locate, with confidence in this measuring point improving markedly.
Chemical & Pharmaceutical · Controllers and Sensors — Application Case
Customer Profile
The typical setting in Chemical & Pharmaceutical is reactors, metering vessels and finished-product tanks. The measuring point is Controllers and Sensors; the focus is online water monitoring with a controller and sensor combination.
Key Difficulties
Conditions are highly corrosive media with solvents and crystallisation, partly in explosive atmospheres and, on top of that, corrosion plus Ex protection shortens instrument life and makes downtime costly, which a standard configuration cannot handle for long. The design therefore starts with matching each controller channel to its sensor type.
Implementation
The scope ran through the Hach channel: supplying against the Hach original part numbers. Technically it hinges on matching each controller channel to its sensor type, and on these requirements: Ex certification and GMP traceability of records.
Customer Benefit
In practice this means factory-matched combinations that make faults easier to locate, which cut the workload on the site maintenance team.
Power & Energy · Controllers and Sensors — Application Case
Project Background
The project covers boiler feedwater, turbine condensate and FGD/DeNOx systems. The work centres on Controllers and Sensors and the core task is online water monitoring with a controller and sensor combination.
Operating Challenges
The medium and duty are characterised by high temperature and pressure with both steam and saline water, and drift under high temperature and pressure, with very costly unplanned outages. So Controllers and Sensors cannot be specified generically — it calls for matching each controller channel to its sensor type.
Solution
The Ainstru engineering team worked through the Hach channel: supplying against the Hach original part numbers. The configuration involves matching each controller channel to its sensor type, with these requirements in view: compliance with power-industry codes and unit availability targets.
Results
Since start-up the system has run steadily; the practical result is factory-matched combinations that make faults easier to locate.