COD-60A Permanganate Index Analyzer
Application industries:
The COD-60A permanganate index analyzer can be used for the determination of oxygen demand / permanganate index in drinking water, surface water, groundwater, seawater, industrial water, and other fields.
Instrument features:
1) The COD-60A permanganate index analyzer uses the accurate coulometric titration method. It responds sensitively and provides accurate data.
2) The digestion time of this permanganate index analyzer is adjustable within a range of approximately 10 to 800 seconds. The permanganate index measurement data can be obtained in about 5 to 10 minutes, making it convenient and quick.
3) Multiple ranges, direct reading of COD values: The range is set according to the dilution ratio, allowing direct reading of the COD value of the undiluted original solution without conversion.
4) Simple operation: The measurement unit is integrated, combining oxidation and titration processing steps. After setting the instrument, one-touch operation automatically completes the entire process from zero calibration, oxidation, and titration to data printing.
5) Convertible to standard method measured values: Calibration and coefficient adjustment can be performed using the standard method. By determining in advance the correlation between the laboratory standard analytical method and the COD-60A permanganate index analyzer, and inputting the linear regression equation, converted values for the standard analytical method can be obtained.
6) Automatic standard program control reduces operating errors: This permanganate index analyzer uses a proprietary system control to automatically perform steps such as oxidation state (heating/cooling) and titration. It automatically controls temperature and, based on coulometric titration, performs titration using self-cleaning electrodes. This reduces human errors between operators.
7) Various maintenance/inspection functions: The COD-60A permanganate index analyzer uses an electrode cleaning mode and a self-check titration mode, making maintenance and inspection work easy.
8) Data storage function (100 sets), built-in printer, and clock: The permanganate index measurement results can be printed after measurement is completed, and historical data can be displayed or printed later.
Measurement method: Acidic (or alkaline) potassium permanganate method
Measurement principle: Coulometric titration (endpoint detection by redox potential method)
Digestion time: Adjustable from 10 to 800 s
Titration temperature: 65℃
Display: Graphic LCD display
During measurement: real-time display of sequence status, sample solution temperature, titration curve, etc.
Error messages: titration failure, blank error, out of effective range
Measurement range: Standard: 0~20/40/100/200/400/1000mg/L
Arbitrary: 0~(10~2000mg/L settable)
Dilution required for 0~20mg/L and above
Repeatability: ±2%FS
Resolution: 0.01mg/L
Program control: Fully automatic measurement processing from oxidation to titration (acidic method only)
Calculation functions: Regression formula conversion function
Statistical calculation (average value)
Automatic input of blank value (or manual input)
Data storage: 100 data records (measurement time, sample number, result)
Printer: Built-in
Output: RS-232C, external printer
Power supply: AC220V50/60Hz
Weight: Main unit: approx. 3.5kg
Related reagents
| Order No. | Description | Parameters | Analysis method | Packaging | Range | Guaranteed shelf life (days) |
|---|---|---|---|---|---|---|
| 2125925-CN | Chemical oxygen demand (COD) digestion reagent, high range, HW, pk/25 | COD (chemical oxygen demand) | 8000 DR800 low range, medium range, high range | 25/pk | 0-150mg/L, COD | 150 |
| 2125825-CN | rr Chemical oxygen demand (COD) digestion reagent, low range, HW, pk/25 | Chemical oxygen demand | Chemical oxygen demand (COD) 8000 3-150mg/L Chemical oxygen demand (COD) 8000 0-150mg/L Chemical oxygen demand (COD) 8000 3-150mg/L low range Chemical oxygen demand (COD) 8000 20-1500mg/L Chemical oxygen demand (COD) 8000 20-1,500mg/L |
25/pk | 3-150mg/L, COD | 150 |
| 2415925-CN | rr Chemical oxygen demand (COD) digestion reagent, high range+, pk/25 | - | Chemical oxygen demand (COD) 8000 200-15,000mg/L Chemical oxygen demand (COD) 8000 0-15,000mg/L |
- | 200-15,000mg/L | 150 |
| TNT820-CN | rr Chemical oxygen demand (COD), TNT+, ultra-low range (1.0-60mg/L), pk/24 | COD (chemical oxygen demand) | 10211 TNTplus ultra-low range | - | 1.0-60.0mg/L | 150 |
| 2415825-CN | rr Chemical oxygen demand (COD) digestion reagent, ultra-low range, pk/25 | COD (chemical oxygen demand) | 8000 digestion colorimetric | - | 0.7-40.0mg/L | 150 |
Ordering guide
Complete unit Order No.: COD60A-2-6
Standard accessories
Name Order No. Quantity
Measuring vessel, support rod, and positioner 0KK00001 1
Electrode sheath 0IB00011 1
Measurement electrode 1 0LA00001 1
Measurement electrode 2 0LA00002 1
Measuring cell 0DA00002 2
Measuring cell holder 0LZ00003 1
Stirring bar 0SE00001 1
Reagent 1 (100ml) 143C143 1
Reagent 2 (25ml) 143C144 1
Electrode filling solution (25ml) 143C145 1
Micropipette (0.5ml) 0SZ00013 1
Liquid measuring pipette (5ml) 0SZ00014 1
Pipette (2ml) 0SZ00015 1
Pipette for filling solution replenishment 0SZ00016 1
User manual 0PA00064 1
Source: Public information from the HACH Water Quality website
Related Products
HACHWater quality analysis test strip series
HACHDigester (rapid COD digester)
HACHMulti-channel electrochemical analyzer
HACHPro/Pro+ pocket water tester
HACHLiquid colorimeter
HACHPortable water toxicity analyzer
HACH1000 series automatic potentiometric titrator
HACH