- Principle of operation:
Total phosphorus measurement (compliant with GB 11893-89): Potassium persulfate is used as the oxidant to digest the water sample at 120°C for 30 min, converting phosphorus compounds into phosphate ions. The total phosphorus content in water is determined by the molybdenum blue spectrophotometric method (measuring wavelength: 700 nm).
Total nitrogen measurement (compliant with standard HJ 636-2012): Potassium persulfate is used as the oxidant to digest the water sample at 120°C for 30 min, converting nitrogen compounds into nitrate ions. The pH of the sample solution is adjusted to 2-3, and the total nitrogen content in water is determined by UV spectrophotometric method (measuring wavelength: 220 nm, 275 nm turbidity correction: A = A220 - A275 × 2).
COD (UV) measurement: Dual-wavelength spectrophotometric method (UV 254 nm, visible light 546 nm).
Measuring range: Total nitrogen, 0~2 mg/L to 200 mg/L
Total phosphorus, 0~0.5 mg/L to 20 mg/L
Repeatability: ≤± 3% FS
Analysis interval: 1 hour~6 hours, freely programmable
Sample conditions: Temperature, 2~40°C
Pressure, 0.02~0.05 MPa
Flow rate, 1~3 L/min
Sample volume per analysis approx. 67.5 mL
Reagent replenishment interval: 1 month
Ambient conditions: Temperature, 2~40°C, 85% relative humidity, indoor installation
Analog output: 4~20 mA, load 600 ohm
Relays: Configurable upper/lower limit alarms, fault messages, etc.
Power supply: 220 VAC, 50/60 HZ
Power consumption: 500 VA (max.) 200 VA (average)
Data communication: RS485 serial port
Dimensions: 450 × 380 × 1430 mm
Weight: approx. 80 Kg
- Application industries:
The NPW-160 total nitrogen analyzer / total phosphorus analyzer is mainly used for automatic monitoring of total phosphorus, total nitrogen, and COD in surface water, tap water, municipal wastewater, and other special applications.
- Instrument features:
● The NPW-160 total nitrogen analyzer / total phosphorus analyzer features a built-in multi-wavelength detector capable of measuring total phosphorus, total nitrogen, and COD (UV).
● Independently designed heating digestion unit, and integrated digestion and measurement design within the total phosphorus / total nitrogen analyzer system, simplifying tubing connections.
● The NPW-160 total nitrogen analyzer / total phosphorus analyzer has a built-in CF card for long-term data storage.
● Equipped with a Chinese touchscreen operation interface for fast and accurate readings, combined with wireless transmission for intelligent and reliable operation.
● Low reagent consumption and waste liquid discharge, low operating and maintenance costs, and minimal secondary pollution.
Ordering guide
Standard configuration includes: main unit, operation manual, installation accessories, etc.
NPW160-0-2C11EA000A001
Total phosphorus: 0~0.5 mg/L (factory setting)
Total nitrogen: 0~20 mg/L (factory setting)
20 mm measuring cell, suitable for surface water, tap water, and compliant municipal wastewater discharge outlets
NPW160-0-2C11GE000A001
Total phosphorus: 0~10 mg/L (factory setting)
Total nitrogen: 0~50 mg/L (factory setting)
10 mm measuring cell, suitable for the wastewater industry
NPW160-0-2C11HF000A001
Total phosphorus: 0~20 mg/L (factory setting)
Total nitrogen: 0~100 mg/L (factory setting)
5 mm measuring cell, suitable for special applications
Recomendados similares
HACHAnalizador de fósforo total / ortofosfato Sigma
Analizador COD en línea por método dicromato sc
HACHAnalizador de dureza del agua en línea
HACHAnalizador automático multiparamétrico de agua
HACHAnalizador multiparamétrico de agua en línea
HACHAnalizador de sílice en línea
HACHAnalizador de nitrógeno amoniacal
HACH