Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

NPW-160 Total Phosphorus/Total Nitrogen Analyzer

NPW-160 Total Phosphorus/Total Nitrogen Analyzer
NPW
HACH NPW
The NPW-160 total phosphorus/total nitrogen analyzer is a two-in-one online monitoring instrument.

- Principle of operation:

Total phosphorus measurement (compliant with GB 11893-89): Potassium persulfate is used as the oxidant to digest the water sample at 120°C for 30 min, converting phosphorus compounds into phosphate ions. The total phosphorus content in water is determined by the molybdenum blue spectrophotometric method (measuring wavelength: 700 nm).

Total nitrogen measurement (compliant with standard HJ 636-2012): Potassium persulfate is used as the oxidant to digest the water sample at 120°C for 30 min, converting nitrogen compounds into nitrate ions. The pH of the sample solution is adjusted to 2-3, and the total nitrogen content in water is determined by UV spectrophotometric method (measuring wavelength: 220 nm, 275 nm turbidity correction: A = A220 - A275 × 2).

COD (UV) measurement: Dual-wavelength spectrophotometric method (UV 254 nm, visible light 546 nm).

Technical specifications
Measuring range:    Total nitrogen, 0~2 mg/L to 200 mg/L
Total phosphorus, 0~0.5 mg/L to 20 mg/L
Repeatability:    ≤± 3% FS
Analysis interval:    1 hour~6 hours, freely programmable
Sample conditions:    Temperature, 2~40°C
Pressure, 0.02~0.05 MPa
Flow rate, 1~3 L/min
Sample volume per analysis approx. 67.5 mL
Reagent replenishment interval:    1 month
Ambient conditions:    Temperature, 2~40°C, 85% relative humidity, indoor installation
Analog output:    4~20 mA, load 600 ohm
Relays:    Configurable upper/lower limit alarms, fault messages, etc.
Power supply:    220 VAC, 50/60 HZ
Power consumption:    500 VA (max.) 200 VA (average)
Data communication:    RS485 serial port
Dimensions:    450 × 380 × 1430 mm
Weight:    approx. 80 Kg

- Application industries:

The NPW-160 total nitrogen analyzer / total phosphorus analyzer is mainly used for automatic monitoring of total phosphorus, total nitrogen, and COD in surface water, tap water, municipal wastewater, and other special applications.

- Instrument features:

● The NPW-160 total nitrogen analyzer / total phosphorus analyzer features a built-in multi-wavelength detector capable of measuring total phosphorus, total nitrogen, and COD (UV).

● Independently designed heating digestion unit, and integrated digestion and measurement design within the total phosphorus / total nitrogen analyzer system, simplifying tubing connections.

● The NPW-160 total nitrogen analyzer / total phosphorus analyzer has a built-in CF card for long-term data storage.

● Equipped with a Chinese touchscreen operation interface for fast and accurate readings, combined with wireless transmission for intelligent and reliable operation.

● Low reagent consumption and waste liquid discharge, low operating and maintenance costs, and minimal secondary pollution.

Ordering guide
Standard configuration includes: main unit, operation manual, installation accessories, etc.
NPW160-0-2C11EA000A001
Total phosphorus:    0~0.5 mg/L (factory setting)
Total nitrogen:    0~20 mg/L (factory setting)
20 mm measuring cell, suitable for surface water, tap water, and compliant municipal wastewater discharge outlets
NPW160-0-2C11GE000A001
Total phosphorus:    0~10 mg/L (factory setting)
Total nitrogen:    0~50 mg/L (factory setting)
10 mm measuring cell, suitable for the wastewater industry
NPW160-0-2C11HF000A001
Total phosphorus:    0~20 mg/L (factory setting)
Total nitrogen:    0~100 mg/L (factory setting)
5 mm measuring cell, suitable for special applications