Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

UVAS Analizador en línea de absorbancia UV sc

UVAS Analizador en línea de absorbancia UV sc
Analizador en línea de absorbancia UV sc
HACH UVAS
UVAS eco sc UV absorption online analyzer is a COD online monitor designed for measuring dissolved organic matter in water (drinking water/surface water/wastewater). It can be directly immersed in water to measure dissolved organic matter in water that absorbs 254 nm UV light, and expresses the measurement results as the "specific absorption coefficient", namely SAC254. The dual-beam structure unique to the measurement probe of this COD online monitor can effectively eliminate interference with measurement results from factors such as turbidity in the sample, power supply fluctuations, and aging of components, thereby improving measurement accuracy. The measurement results can be displayed in graphical or digital form. This COD online monitor features automatic cleaning and requires no chemical reagents. Under suitable conditions, it is the choice for controlling water treatment process parameters.

Typical Applications

  • This online COD analyzer can be used in small and medium-sized wastewater treatment plants that need to monitor organic concentration trends, for continuous monitoring of dissolved organic pollutants in water.

  • Water quality monitoring at the inlet and outlet of wastewater treatment plants, providing users with continuous and stable measurement data at low operating costs.

  • In wastewater treatment processes, it is necessary to control the balance between organic load, aeration rate, and nutrients. Using UVAS eco sc

  • The UV absorption online analyzer can provide continuous organic matter trend data, helping operators optimize the process.

Instrument Features

  • This online COD analyzer has low operating costs, with only the sealing ring and wiper blade as consumable parts.

  • Economical product of UVAS, at a lower price.

  • Stainless steel material, sturdy and durable.

  • Integrated automatic wiper cleaning function, ensuring low maintenance requirements even under very demanding application conditions, with no additional installation and maintenance required (e.g., compressor).

  • Plug and play.

Principle of Operation

When organic substances containing conjugated double bonds or polycyclic aromatic hydrocarbons are dissolved in water, they absorb ultraviolet light. Therefore, by measuring the degree of absorption of 254 nm UV light by these organic substances, we can assess the degree of organic pollution in the water body. The “specific absorption coefficient”, namely SAC254, is the physical quantity used to measure the total amount of organic pollutants in water. This online COD analyzer consists of a controller and a measuring probe. When the measuring probe is in operation, it needs to be immersed in water, or water is pumped up to flow through the gap. When the light emitted by the light source in the probe passes through the gap, part of the light is absorbed by the sample flowing through the gap, while the rest of the light passes through the sample and reaches the beam splitter on the other side of the probe, where it is split into two beams: 50% of the light is detected by the sample detector, and the other 50% is detected by the reference detector. This online COD analyzer compares the signals from the two detectors to provide the “specific absorption coefficient”, namely SAC254.

Technical Specifications

  • Measuring technique:                                  UV absorption measurement (dual-beam technology), no chemical reagents required

  • Resolution:                                      0.5 m-1 or mg/L or ppm

  • Accuracy:                                      ±5% of measured value, ±10 m-1, whichever is greater

  • Measuring optical path:                                  1mm

  • Measuring range:                                       5-1500 m-1

  • Compensation wavelength:                                  550nm

  • Response time:                                  ≥5 minutes, or longer

  • Cable length:                                  10m, extendable to 100m

  • Analog output:                                  Two 0/4~20mA outputs (via sc200)

  • Probe pressure limit                              Max. 0.5bar

  • Ambient sample temperature                              +2~+40℃

  • Water sample pH:                                   4.5~9

  • Dimensions:                                        Approx. 75×323mm

  • Weight:                                        Approx. 3.3kg

  • Service interval:                                 6 months

  • User maintenance:                                Typically 1 hour/month

Ordering Guide

  • Standard configuration includes: UVAS eco sc sensor, sc200 or sc1000 controller, sensor mounting accessories

  • LXV443.80.10002                        UVAS eco sc probe, 1mm optical path

  • LZX414.00.10000                        Sensor mounting accessories

  • sc200 Controller (must be ordered separately)

  • LXV404.99.00552                       Standard sc200 controller, digital probe, dual channel, 220V

  • LXV404.99.01552                       sc200 controller with RS232/RS485 MODBUS network card, digital probe, dual channel, 220V

  • sc1000 Controller (must be ordered separately)

  • LXV402.99.00002                       sc1000 display module

  • LXV400.99.1R572                       sc1000 probe module, 4 sensors, 4mA input, 4mA output, 4 relays, 110-230V

  • LXV400.99.1B572                       sc1000 probe module, 4 sensors, 4mA input, 4mA output, 4 relays, RS485(MODBUS), 110-230V

  • LXV400.99.1F572                       sc1000 probe module, 4 sensors, 4mA input, 4mA output, 4 relays, PROFIBUS DP, 110-230V

  • LXV400.99.1R582                       sc1000 probe module, 6 sensors, 4mA input, 4mA output, 4 relays, 110-230V

Source: Public information from HACH Water Quality website

Tratamiento de Aguas Residuales Municipales

Tratamiento de Aguas Residuales Municipales — Controladores y sensores: caso de aplicación

Situación en planta

Se trata de un cliente del sector Tratamiento de Aguas Residuales Municipales cuya planta está en las etapas de entrada, biológica y salida de una planta municipal. La tarea principal era monitorización en línea con combinación de controlador y sensores.

El problema

La dificultad está en alto contenido de sólidos y componentes corrosivos, con grandes variaciones diarias de caudal; a la vez, los parámetros del efluente oscilan y las sondas se ensucian con lodos y biopelícula. Por eso la especificación implica emparejar cada canal del controlador con su tipo de sensor.

Qué hicimos

Adoptamos un enfoque de Hach: suministrar según referencias originales de Hach. La entrega se construyó en torno a emparejar cada canal del controlador con su tipo de sensor; se verificó punto por punto: cumplimiento estable del nivel 1A y reporte de datos en línea.

Resultado operativo

En servicio el sistema responde a lo previsto: combinaciones de fábrica que facilitan localizar fallos, con un mantenimiento rutinario notablemente menor.

Alimentación y Bebidas

Alimentación y Bebidas — Controladores y sensores: caso de aplicación

Contexto de aplicación

las etapas de dosificación, pasteurización, llenado y limpieza CIP es un servicio típico de Alimentación y Bebidas. El proyecto partió de Controladores y sensores, centrándose en monitorización en línea con combinación de controlador y sensores.

Limitaciones del proceso

las limpiezas frecuentes exigen instrumentos resistentes a ciclos CIP/SIP con sosa caliente, y jarabes, lácteos y bebidas proclives a incrustaciones, con estrictos requisitos higiénicos no hace sino acentuarlo. Por eso la especificación de Controladores y sensores debe cubrir emparejar cada canal del controlador con su tipo de sensor.

Enfoque de suministro

La entrega se gestionó como Hach: suministrar según referencias originales de Hach. La ejecución se centró en emparejar cada canal del controlador con su tipo de sensor, junto con requisitos de seguridad alimentaria y diseño higiénico 3-A / EHEDG.

Resultados en campo

La operación evidencia el beneficio: combinaciones de fábrica que facilitan localizar fallos, con una confianza claramente mayor en este punto de medición.

Química y Farmacéutica

Química y Farmacéutica — Controladores y sensores: caso de aplicación

Perfil del cliente

El entorno típico en Química y Farmacéutica es reactores, vasijas de dosificación y tanques de producto terminado. El punto de medición es Controladores y sensores; el foco es monitorización en línea con combinación de controlador y sensores.

Dificultades clave

Las condiciones son medios muy corrosivos con disolventes y cristalización, en parte en atmósferas explosivas y, encima, la corrosión y la protección Ex acortan la vida útil y encarecen las paradas, algo que una configuración estándar no soporta mucho tiempo. El diseño parte de emparejar cada canal del controlador con su tipo de sensor.

Implementación

El alcance se canalizó por la vía de Hach: suministrar según referencias originales de Hach. Técnicamente se apoya en emparejar cada canal del controlador con su tipo de sensor, y en estos requisitos: certificación Ex y trazabilidad de datos según GMP.

Beneficio para el cliente

En la práctica esto significa combinaciones de fábrica que facilitan localizar fallos, lo que redujo la carga del equipo de mantenimiento.

Energía y Electricidad

Energía y Electricidad — Controladores y sensores: caso de aplicación

Antecedentes del proyecto

El proyecto abarca agua de alimentación, condensado de turbina y sistemas de desulfuración/DeNOx. El trabajo se centra en Controladores y sensores y la tarea principal es monitorización en línea con combinación de controlador y sensores.

Condiciones de servicio

El medio y el servicio se caracterizan por alta temperatura y presión, con vapor y agua salina, y además deriva a alta temperatura y presión, con paradas no programadas muy costosas. Por eso Controladores y sensores no puede especificarse de forma genérica: requiere emparejar cada canal del controlador con su tipo de sensor.

Solución

El equipo técnico de Ainstru trabajó por la vía de Hach: suministrar según referencias originales de Hach. La configuración incluye emparejar cada canal del controlador con su tipo de sensor; se cuidaron estos requisitos: cumplimiento de normativas del sector y objetivos de disponibilidad.

Resultados

Desde la puesta en marcha el sistema funciona de forma estable; el resultado práctico es combinaciones de fábrica que facilitan localizar fallos.