Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

TU5400sc Turbidimeter + SC4500 Controller

TU5400sc Turbidimeter + SC4500 Controller
TU5400sc Turbidimeter + SC4500 Controller
HACH TU5400 SC4500
Configurable Models 2 models
TU5400sc Turbidimeter + SC4500 Controller

产品概述

The next standard in the evolution of turbidity

Compliant with EPA approved Hach Method 10258 and perfect for membrane filtration applications, the TU5400sc employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability between measurements. For the first time you will be able to remove uncertainty of which measurement to trust between your lab and process instruments, thanks to identical 360° x 90° Detection® Technology in any TU5 Series® turbidimeter. The TU5400sc Ultra-High Precision Low Range Laser Turbidimeter dramatically reduces the time needed to get a turbidity measurement you can rely on. With a 10mL sample size, response time is over 15 times faster than an FT660. A stable laser light source eliminates the need for annual lamp replacements. The turbidimeter’s measurement surface area has been reduced by 98%, allowing you to save time cleaning. With a mounting pattern identical to the FT660, replacing your previous turbidimeter is fast and simple. All of this, along with the ability to measure to 0.0002 NTU, gives you the next standard in the evolution of turbidity. A flow meter is included with this turbidimeter that continuously measures flow through the instrument and displays the flow value on screen. Current flow can be output to SCADA via 4-20mA or digital communications. Ensure that your turbidimeter is always functioning properly with this TU5's System Check capability. With the ability to provide information on measurement status and vial cleanliness, you can have full confidence in your turbidity measurement. RFID technology included with this turbidimeter delivers easy calibration and simple, paperless turbidity measurement comparisons between your process turbidimeter and the TU5200.

主要特点

  • Groundbreaking 360° x 90° Detection™ Technology:The TU5 Series employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability from test to test.
  • Everything about turbidity – faster:The TU5 Series dramatically reduces the time needed to get a turbidity measurement you can rely on, with 98% less online sample surface area to clean, sealed vials for calibration, and the elimination of the need for indexing and silicone oil in the lab. Not to mention, a smaller online sample volume means you will detect events almost immediately.
  • Easy Adoption:The familiar experience of a modern touchscreen, the ability to use your current Hach sensors, and the same footprint as the SC200, make installation and integration of the SC4500 Controller seamless.
  • No Time for Downtime:The SC4500's built-in predictive diagnostic software ensures measurement confidence and reduces the risk of unexpected equipment downtime by enabling proactive maintenance planning via MSM, including step-by-step instructions.

技术参数

参数名称说明
Accuracy±2% or 0.01 NTU from 0 - 40 NTU ±10% of reading from 40 - 1000 NTU based on Formazin primary standard
Cable Length1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration MethodFor Formazin and Stablcal®: 20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range Custom calibration for up to 6-point calibrations
CertificationsCE compliant US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version Australian ACMA Marking
CommunicationSystem Check & RFID
Controller2 Channels, 100-240 VAC
Dimensions (H x W x D)249 mm x 268 mm x 190 mm
Enclosure RatingElectronic compartment IP55; all other functional units IP65 with process head/ACM attached to the TU5300sc/TU5400sc instrument
Fitting TypeSample quick connector: ¼-in. for ¼-in. tubing
InstrumentWith Flow Sensor
Light SourceClass 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
MaterialASA Luran S 777K / RAL7000, TPE RESIN Elastocon® STK40, Thermoplastic Elastomer TPS-SEBS
ModelTU5400sc
Mounting ConfigurationsWall Mount
Operating HumidityRelative humidity: 5 - 95% at different temperatures, non-condensing
Operating Temperature Range0 - 50 °C (32 - 122 °F)
OptionsProcess Head & Flow Sensor
RangeEPA: 0 - 700 NTU / FNU / TE/F / FTU 0 - 100 mg/L 0 - 175 EBC ISO: 0 - 1000 NTU / FNU / TE/F / FTU 0 - 100 mg/L 0 - 250 EBC
RegulatoryEPA
RepeatabilityBetter than 1% of reading or ±0.002 NTU (TU5300sc) or ±0.0006 NTU (TU5400sc) on Formazin at 25 °C (77 °F), whichever is greater
Resolution0.0001 NTU / FNU / TE/F / FTU / EBC
Response TimeTU5300sc: T90 <45 seconds at 100 mL/min TU5400sc: T90 <30 seconds at 100 mL/min
Sample Flow Rate100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
Sample Pressure6 bar (87 psi) maximum, compared to air at sample temperature range from 2 - 40 °C (35.6 - 104 °F)
Sample Temperature2 - 60 °C (35 - 140 °F)
Signal Average TimeTU5300sc: 30 - 90 seconds TU5400sc: 1 - 90 seconds
Storage Conditions-40 - 60 °C (-40 - 140 °F)
Stray Light<10 mNTU
UnitsNTU, FNU, TE/F, FTU, EBC
VerificationLiquid: Stablcal®, Formazin (0.1 - 40 NTU) Dry: Glass Rod at <0.1 NTU
Weight5.95 lbs. (2.7 kg); 11 lbs. (5.0 kg) with all accessories
What's included?Controller, Non-Claros + Embedded Prognosys, without power cord; includes mounting hardware

应用领域

Filtration · Membrane Protection · Regulations · Water Reuse

同系列其他配置

型号
LXV525.97KTO039TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, EPA, Auto

资料下载

2 models · click to collapse
Municipal Wastewater Treatment

Municipal Wastewater Treatment · Controllers and Sensors — Application Case

Site Conditions

This is a Municipal Wastewater Treatment customer whose plant is at the inlet, biological and outlet stages of a municipal plant. The main task was online water monitoring with a controller and sensor combination.

The Problem

The difficulty lies in high solids and corrosive fractions, with large diurnal flow swings; at the same time effluent parameters drift and probes foul with sludge and biofilm. Specification therefore means matching each controller channel to its sensor type.

What We Did

We took a Hach channel approach — supplying against the Hach original part numbers. Delivery was built around matching each controller channel to its sensor type, with point-by-point checks on: stable compliance with Grade 1A discharge limits and online data reporting.

Operating Outcome

In service the system performs as intended — factory-matched combinations that make faults easier to locate, with routine maintenance falling noticeably.

Food & Beverage

Food & Beverage · Controllers and Sensors — Application Case

Application Context

dosing, pasteurising, filling and CIP stages is a typical Food & Beverage duty. The project started from Controllers and Sensors, focusing on online water monitoring with a controller and sensor combination.

Process Constraints

frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles, and syrups, dairy and beverages that scale easily, under strict hygienic requirements only amplify it. The Controllers and Sensors specification must therefore cover matching each controller channel to its sensor type.

Delivery Approach

Delivery was handled as the Hach channel — supplying against the Hach original part numbers. Execution centred on matching each controller channel to its sensor type, together with food-safety and hygienic design per 3-A / EHEDG.

Field Results

Operation shows the benefit clearly — factory-matched combinations that make faults easier to locate, with confidence in this measuring point improving markedly.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · Controllers and Sensors — Application Case

Customer Profile

The typical setting in Chemical & Pharmaceutical is reactors, metering vessels and finished-product tanks. The measuring point is Controllers and Sensors; the focus is online water monitoring with a controller and sensor combination.

Key Difficulties

Conditions are highly corrosive media with solvents and crystallisation, partly in explosive atmospheres and, on top of that, corrosion plus Ex protection shortens instrument life and makes downtime costly, which a standard configuration cannot handle for long. The design therefore starts with matching each controller channel to its sensor type.

Implementation

The scope ran through the Hach channel: supplying against the Hach original part numbers. Technically it hinges on matching each controller channel to its sensor type, and on these requirements: Ex certification and GMP traceability of records.

Customer Benefit

In practice this means factory-matched combinations that make faults easier to locate, which cut the workload on the site maintenance team.

Power & Energy

Power & Energy · Controllers and Sensors — Application Case

Project Background

The project covers boiler feedwater, turbine condensate and FGD/DeNOx systems. The work centres on Controllers and Sensors and the core task is online water monitoring with a controller and sensor combination.

Operating Challenges

The medium and duty are characterised by high temperature and pressure with both steam and saline water, and drift under high temperature and pressure, with very costly unplanned outages. So Controllers and Sensors cannot be specified generically — it calls for matching each controller channel to its sensor type.

Solution

The Ainstru engineering team worked through the Hach channel: supplying against the Hach original part numbers. The configuration involves matching each controller channel to its sensor type, with these requirements in view: compliance with power-industry codes and unit availability targets.

Results

Since start-up the system has run steadily; the practical result is factory-matched combinations that make faults easier to locate.