Technical Data
CompareConductivity
| Measuring Principle | |
| Application | Phase seperation, chromotography, fermentation, CIP monitoring in small pipes, ultrafiltration |
| Characteristic | Conductivity sensor with a broad measurement range for all hygienic applications with advanced requirements to process safety |
| Measuring range | 1 µS/cm to 500 mS/cm |
| Measuring principle | 4-electrode conductive conductivity sensor |
| Design | Certified hygienic design with FDA compliant materials, available as 120 mm sensor with PG13.5 and in many other standard process connections |
| Material | Process connection: stainless steel 1.4435 (316L) Sensor element: platinum and ceramic |
| Dimension | Sensor diameter: 12 mm (0.47 inch) Electrode length: depending on process connection |
| Process temperature | -5 to 120 °C (23 to 248 °F) Sterilization: max. 140 °C at 6 bar for max. 45 min (Max. 284 °F at 87 psi for max. 45 min) |
| Process pressure | 17 bar abs at 20 °C (247 psi at 68 °F) 9 bar abs at 120 °C (131 psi at 248 °F) |
| Temperature sensor | Pt1000 |
| Connection | Clamp 1,5", Clamp 2", Varivent F DN25,Varivent N DN40-DN125, BioControl DN25, PG13,5, DN25 standard/40mm, DN25 B.Braun port/65mm, Threat G1", Thread NPT 1" Cable: Memosens connector |
| Ingress protection | IP 68 / NEMA type 6P |
| Certificates, declarations and approvals | 3-A, hygienic design EHEDG, ASME BPE-2002 Pharma CoC, FDA, Regulation (EC) No. 1935/2004 |
The Memosens CLS82D is a hygienic digital sensor for reliable, safe and fast measurement in applications with strongly fluctuating conductivity values. Multiple certificates, various process connections and a small compact design meet your process requirements. Using Memosens digital technology, the CLS82D is easy to operate and offers the highest process safety and data integrity. Moisture- and corrosion-resistant, it allows laboratory calibration and device predictive maintenance.
Advantages
Wide measuring range
Hygienic design approved by EHEDG and 3-A
Memosens technology and unique electrode connection monitoring ensure absolute loop safety
Steam sterilization and high-temperature sterilization
Exact electrode constant indicated in the quality certificate
Sensor parameters are recorded for simple traceability and predictive maintenance
Convenient maintenance strategy with easy calibration and sensor management using Memobase Plus
Areas of application
The Memosens CLS82D can measure over a wide conductivity range:
Phase separation
Downstream process stages
Fermenters
Final rinse in pharmaceutical applications
International explosion protection approvals such as ATEX, FM, CSA and NEPSI; the digital sensor is suitable for use in hazardous areas.
Spare Parts & Accessories 12
12 spare parts · click to collapse
| Order code | Description | |
|---|---|---|
| 71307105 | CLS82D Accessory: Fluoroelastomer GasketThe kit can be used for the sensors CLS82D/E-**NA* and CLS82D/E-**NB* | Details |
| 71307106 | CLS82D Accessory: EPDM GasketThe kit can be used for the sensors CLS82D/E-**NA* and CLS82D/E-**NB* | Details |
| 71307107 | CLS82D Accessory: Silicone GasketThe kit can be used for the sensors CLS82D/E-**NA* and CLS82D/E-**NB* | Details |
| 51508049 | Welding stud CPA442, straight, 50mmMaterial 1.4435, DN25-G1 1/4" for CPA442-B | Details |
| 51508050 | Welding stud CPA442, oblique, 50/60 mmMaterial 1.4435, DN25-G1 1/4" for CPA442-B | Details |
| 51508051 | Welding stud CPA442, straight, 30mmMaterial 1.4435, DN25-G1 1/4" for CPA442-A | Details |
| 51508052 | Welding stud CPA442, oblique, 30/40 mmMaterial 1.4435, DN25-G1 1/4" for CPA442-A | Details |
| 50081902 | Calibration solution 74.0 µS/cm500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO | Details |
| 50081903 | calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO. | Details |
| 50081904 | Calibration solution 1.406 mS/cm500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO | Details |
| 50081905 | calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM to NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO | Details |
| 50081906 | calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO | Details |
Configurator / Comparison
A
Source:Endress+Hauserwebsite public information
Measuring principle video
Click to playManuals & Documents 5
All documents for this modelCase Studies
Municipal Wastewater Treatment · conductivity — Application Case
Customer Profile
The typical setting in Municipal Wastewater Treatment is the inlet, biological and outlet stages of a municipal plant. The measuring point is conductivity; the focus is measuring conductivity to judge water purity or chemical concentration.
Key Difficulties
Conditions are high solids and corrosive fractions, with large diurnal flow swings and, on top of that, effluent parameters drift and probes foul with sludge and biofilm, which a standard configuration cannot handle for long. The design therefore starts with choosing two- or four-electrode cells by range and corrosivity.
Implementation
The scope ran through the analyser range: selecting as part of a complete water-analysis system. Technically it hinges on choosing two- or four-electrode cells by range and corrosivity, and on these requirements: stable compliance with Grade 1A discharge limits and online data reporting.
Customer Benefit
In practice this means stable concentration control and less reagent and discharge waste, which cut the workload on the site maintenance team.
Food & Beverage · conductivity — Application Case
Project Background
The project covers dosing, pasteurising, filling and CIP stages. The work centres on conductivity and the core task is measuring conductivity to judge water purity or chemical concentration.
Operating Challenges
The medium and duty are characterised by syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. So conductivity cannot be specified generically — it calls for choosing two- or four-electrode cells by range and corrosivity.
Solution
The Ainstru engineering team worked through the analyser range: selecting as part of a complete water-analysis system. The configuration involves choosing two- or four-electrode cells by range and corrosivity, with these requirements in view: food-safety and hygienic design per 3-A / EHEDG.
Results
Since start-up the system has run steadily; the practical result is stable concentration control and less reagent and discharge waste.
Chemical & Pharmaceutical · conductivity — Application Case
Project Overview
The site is reactors, metering vessels and finished-product tanks. The customer wanted conductivity properly configured on the existing plant, with the aim of measuring conductivity to judge water purity or chemical concentration.
Technical Challenges
What actually causes trouble is highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. All of this means the conductivity configuration has to involve choosing two- or four-electrode cells by range and corrosivity.
Configuration
Supply and installation were organised through the analyser range: selecting as part of a complete water-analysis system. On site this meant choosing two- or four-electrode cells by range and corrosivity, checked line by line against: Ex certification and GMP traceability of records.
Performance
After commissioning the outcome is stable concentration control and less reagent and discharge waste, and the readings are now used for process tuning.
Power & Energy · conductivity — Application Case
Site Conditions
This is a Power & Energy customer whose plant is at boiler feedwater, turbine condensate and FGD/DeNOx systems. The main task was measuring conductivity to judge water purity or chemical concentration.
The Problem
The difficulty lies in high temperature and pressure with both steam and saline water; at the same time drift under high temperature and pressure, with very costly unplanned outages. Specification therefore means choosing two- or four-electrode cells by range and corrosivity.
What We Did
We took a analyser range approach — selecting as part of a complete water-analysis system. Delivery was built around choosing two- or four-electrode cells by range and corrosivity, with point-by-point checks on: compliance with power-industry codes and unit availability targets.
Operating Outcome
In service the system performs as intended — stable concentration control and less reagent and discharge waste, with routine maintenance falling noticeably.