Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

CLS82E Memosens

CLS82E Memosens
Memosens CLS82E
Configurable Models 5 models
Memosens CLS82E is a high-end, hygienic sensor for applications with widely varying conductivity values. It enables reliable monitoring of core processes as well as final rinse with only one sensor. FDA compliance and a small, compact design ensure the perfect fit to your processes. With Memosens 2.0 digital technology, CLS82E allows for extended storage of process and sensor data facilitating lab calibration and predictive maintenance and providing the perfect basis for enhanced IIoT services.
Compare

Conductivity

Measuring Principle Conductive
Application Phase seperation, chromotography, fermentation
CIP monitoring in small pipes
Ultrafiltration
Final rinse
Characteristic Conductivity sensor with a broad measurement range for all hygienic applications with advanced requirements to process safety
Measuring range 1 µS/cm to 500 mS/cm
Measuring principle Hygienic 4-electrode conductivity sensor
Design Certified hygienic design with FDA compliant materials, 3-A and EHEDG certified, available as 120 mm sensor with PG13.5 and in many other standard process connections
Material Process connection: stainless steel 1.4435 (316L)
Sensor element: platinum and ceramic
Dimension Sensor diameter: 12 mm (0.47 inch)
Electrode length: depending on process connection
Process temperature -5 to 120 °C (23 to 248 °F)
Sterilization: max. 140 °C at 6 bar for max. 45 min
(Max. 284 °F at 87 psi for max. 45 min)
Process pressure 17 bar abs at 20 °C (247 psi at 68 °F)
9 bar abs at 120 °C (131 psi at 248 °F)
Temperature sensor Pt1000
Ex certification ATEX, NEPSI, CSA, IECE, EAC Ex, INMETRO
Connection Clamp 1,5", Clamp 2", Varivent F DN25,Varivent N DN40-DN125, BioControl DN25, PG13,5, DN25 standard/40mm, DN25 B.Braun port/65mm, Threat G1", Thread NPT 1"
Sensor connection: Inductive, digital connection head with Memosens 2.0 technology
Ingress protection IP 68 / NEMA Type 6P
Certificates, declarations and approvals Inspection certificate 3.1, EN10204
Surface roughness test ISO4287/Ra (≤ 0,38 µm)
Conformity to cGMP derived requirements
ASME BPE
US Food Contact Materials FDA CFR 21
EU Food Contact Materials (EG) 1935/2004
CN Food Contact Materials GB 4806
EHEDG EL Class I, 3-A

Memosens CLS82E is a high-end hygienic sensor suitable for applications requiring a wide conductivity range. A single sensor reliably monitors critical process steps and the final rinsing step. It is FDA-certified and its small, compact design adapts perfectly to the actual process. CLS82E uses Memosens 2.0 digital technology, supports extended data storage, saves calibration and process parameters and allows laboratory pre-calibration of the sensor, providing the perfect basis for implementing predictive maintenance and advancing IIoT technology.

Advantages
Hygienic design with material certifications, EHEDG tested, 3-A and FDA approvals ensure GMP compliance. The sensor withstands high-temperature disinfection and autoclaving.
With Memosens technology, the electrode features a dedicated buildup detection function to keep the measuring loop safe at all times.
Wide measuring range; a single sensor reliably monitors critical process steps and the final rinsing step, saving costs.
The sensor is small and compact and can be installed in small-diameter pipes and confined spaces for measurement.
The quality certificate indicates the electrode's cell constant, precisely adjusted to obtain high-precision measurement results.
Non-contact inductive signal transmission provides high process and data integrity.
Supports IIoT: Memosens 2.0 supports extended data storage, saves calibration and process parameters for better trend identification, and is the perfect basis for implementing predictive maintenance and advancing IIoT technology.

Application
The Memosens CLS82E conductivity sensor has a wide range of applications:
Phase separation
Downstream processing
Fermentation
Chromatography
Ultrafiltration
Final rinsing step in pharmaceutical production
With international explosion protection approvals such as ATEX, FM, CSA and NEPSI, the digital sensor can be used in hazardous areas.

 

Spare Parts & Accessories 12

12 spare parts · click to collapse
Order codeDescription
51508049Welding stud CPA442, straight, 50mmMaterial 1.4435, DN25-G1 1/4" for CPA442-BDetails
51508050Welding stud CPA442, oblique, 50/60 mmMaterial 1.4435, DN25-G1 1/4" for CPA442-BDetails
51508051Welding stud CPA442, straight, 30mmMaterial 1.4435, DN25-G1 1/4" for CPA442-ADetails
51508052Welding stud CPA442, oblique, 30/40 mmMaterial 1.4435, DN25-G1 1/4" for CPA442-ADetails
50081902Calibration solution 74.0 µS/cm500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISODetails
50081903calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO.Details
50081904Calibration solution 1.406 mS/cm500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISODetails
50081905calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM to NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISODetails
50081906calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISODetails
71307105CLS82D Accessory: Fluoroelastomer GasketThe kit can be used for the sensors CLS82D/E-**NA* and CLS82D/E-**NB*Details
71307106CLS82D Accessory: EPDM GasketThe kit can be used for the sensors CLS82D/E-**NA* and CLS82D/E-**NB*Details
71307107CLS82D Accessory: Silicone GasketThe kit can be used for the sensors CLS82D/E-**NA* and CLS82D/E-**NB*Details

Configurator / Comparison

ConfigureEnter a product model to configure
CompareEnter the product models to compare — up to three products

 

 

A

Source:Endress+Hauserwebsite public information

5 models · click to collapse
All documents for this model
4 documents · click to expand
BA

Memosens CLS82E Chinese

Rev. 0321 20 pages
Preview
BA

Memosens CLS82E English

Rev. 0321 20 pages
Preview
TI

TI Memosens CLS82E Chinese

Rev. 0324 16 pages
Preview
TI

TI Memosens CLS82E English

Rev. 0324 16 pages
Preview
Municipal Wastewater Treatment

Municipal Wastewater Treatment · conductivity — Application Case

Customer Profile

The typical setting in Municipal Wastewater Treatment is the inlet, biological and outlet stages of a municipal plant. The measuring point is conductivity; the focus is measuring conductivity to judge water purity or chemical concentration.

Key Difficulties

Conditions are high solids and corrosive fractions, with large diurnal flow swings and, on top of that, effluent parameters drift and probes foul with sludge and biofilm, which a standard configuration cannot handle for long. The design therefore starts with choosing two- or four-electrode cells by range and corrosivity.

Implementation

The scope ran through the analyser range: selecting as part of a complete water-analysis system. Technically it hinges on choosing two- or four-electrode cells by range and corrosivity, and on these requirements: stable compliance with Grade 1A discharge limits and online data reporting.

Customer Benefit

In practice this means stable concentration control and less reagent and discharge waste, which cut the workload on the site maintenance team.

Food & Beverage

Food & Beverage · conductivity — Application Case

Project Background

The project covers dosing, pasteurising, filling and CIP stages. The work centres on conductivity and the core task is measuring conductivity to judge water purity or chemical concentration.

Operating Challenges

The medium and duty are characterised by syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. So conductivity cannot be specified generically — it calls for choosing two- or four-electrode cells by range and corrosivity.

Solution

The Ainstru engineering team worked through the analyser range: selecting as part of a complete water-analysis system. The configuration involves choosing two- or four-electrode cells by range and corrosivity, with these requirements in view: food-safety and hygienic design per 3-A / EHEDG.

Results

Since start-up the system has run steadily; the practical result is stable concentration control and less reagent and discharge waste.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · conductivity — Application Case

Project Overview

The site is reactors, metering vessels and finished-product tanks. The customer wanted conductivity properly configured on the existing plant, with the aim of measuring conductivity to judge water purity or chemical concentration.

Technical Challenges

What actually causes trouble is highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. All of this means the conductivity configuration has to involve choosing two- or four-electrode cells by range and corrosivity.

Configuration

Supply and installation were organised through the analyser range: selecting as part of a complete water-analysis system. On site this meant choosing two- or four-electrode cells by range and corrosivity, checked line by line against: Ex certification and GMP traceability of records.

Performance

After commissioning the outcome is stable concentration control and less reagent and discharge waste, and the readings are now used for process tuning.

Power & Energy

Power & Energy · conductivity — Application Case

Site Conditions

This is a Power & Energy customer whose plant is at boiler feedwater, turbine condensate and FGD/DeNOx systems. The main task was measuring conductivity to judge water purity or chemical concentration.

The Problem

The difficulty lies in high temperature and pressure with both steam and saline water; at the same time drift under high temperature and pressure, with very costly unplanned outages. Specification therefore means choosing two- or four-electrode cells by range and corrosivity.

What We Did

We took a analyser range approach — selecting as part of a complete water-analysis system. Delivery was built around choosing two- or four-electrode cells by range and corrosivity, with point-by-point checks on: compliance with power-industry codes and unit availability targets.

Operating Outcome

In service the system performs as intended — stable concentration control and less reagent and discharge waste, with routine maintenance falling noticeably.