The FQG61 source container is designed to hold the radioactive source during radiometric point level detection, continuous level and density measurement. The radiation is emitted almost unattenuated in one direction only, and is damped in all other directions. This guarantees highest safety for the personnel and a reliable measurement.
Compare

Continuous / Liquids

Measuring Principle Radiometric
Characteristic / Application Source container
Emission angle: 40 / 20 degrees
Approximately 40kg
Specialities Manual or pneumatic switch-on/ switch-off
Ambient temperature -40 °C...+200 °C
(-40 °F...+392 °F)
Process temperature Any
Process pressure / max. overpressure limit Any
Main wetted parts Non-contact
Process connection Non-contact
Certificates / Approvals ATEX, GOST

Continuous / Solids

Measuring Principle Radiometric
Characteristic / Application Source container
Emission angle: 40 / 20 degrees
Approximately 40kg
Specialities Control area calculation with Applicator
Ambient temperature -40 °C...+200 °C
(-40 °F...+392 °F)
Process temperature Any
Process pressure / max. overpressure limit Any
Main wetted parts Non-contact
Process connection Non-contact
Certificates / Approvals ATEX, GOST

Density

Measuring Principle Radiometric Density
Characteristic / Application Source container
Emission angle: 5/ 20/ 40 degrees
Approximately 40kg
Ambient temperature -40 °C...+200 °C
(-40 °F...+392 °F)
Process temperature Any
Process pressure Any
Wetted parts Non-contact
Hygienic Non-contact
Specialities Control area calculation with Applicator

Point Level / Liquids

Measuring Principle Radiometric Limit
Characteristic / Application Source container
Emission angle: 5 degrees
Approximately 40kg
Specialities Control area calculation with Applicator
Ambient temperature -40 °C...+200 °C
(-40 °F...+392 °F)
Process temperature Any
Process pressure / max. overpressure limit Any
Main wetted parts Non-contact
Process connection Non-contact
Certificates / Approvals ATEX, GOST

Point Level / Solids

Measuring Principle Radiometric Limit
Characteristic / Application Source container
Emission angle: 5 degrees
Approximately 40kg
Specialities Control area calculation with Applicator
Ambient temperature -40 °C...+200 °C
(-40 °F...+392 °F)
Process temperature Any
Process pressure / max. overpressure limit Any
Main wetted parts Non-contact
Process connection Non-contact
Certificates / Approvals ATEX, GOST

The FQG61 source capsule contains the radiation source and is used for radiometric limit detection as well as continuous level and density measurement. The radiation is emitted almost without attenuation in only one direction and is shielded in all other directions, ensuring maximum personnel safety and reliable measurement.

Benefits

  • Lightweight source capsule with near-spherical design for excellent shielding

  • Safe and convenient source replacement.

  • Meets the highest safety class (DIN 25426/ISO 2919, typical category C66646).

  • One-piece source capsule, simple installation.

  • Various radiation angles for optimum application results

  • Manual or pneumatic opening and closing

  • Switch position fixed by padlock, cylinder lock or safety bolt

Applications

  • Used to house the radiation source during radiometric limit detection, continuous level measurement and density measurement. The radiation is unattenuated in only one direction and damped in all other directions.

  • Activity: 137Cs... 18.5 GBq (500 mCi), 60Co ... 0.74 GBq (20 mCi)

  • Temperature: unrestricted (non-immersed, external measurement)

  • Pressure: unrestricted (non-immersed, external measurement)

Spare Parts & Accessories 11

11 spare parts · click to collapse
Order codeDescription
71099266Hood FQG61/FQG62, chain, SteelUsage: FQG61/FQG62_020_A.Details
71099267maskUsage: FQG61/FQG62.Details
71301302padlockWith key, simultaneous locking. Usage: FQG61/FQG62/FQG66.Details
52018920keySource holder QG020/QG100Details
71227681keyUsage: cylinder lock, FQG61, FQG62, FQG74.Details
71308587Junction BoxWith fixing screws. Usage: FQG61/FQG62.Details
71099421nameplateWhen changing the source. Consists of: 1x label "radioactive". 3x nameplate, blank. Grooved drive stud, seal wire. Usage: FQG60/FQG61/FQG62/FQG63.Details
71383377ScrewKnurled screw 6x41x15mm. Material: 1.4404. Usage: FQG61/FQG62.Details
71383378ScrewGrub screw ISO4027 M6. Usage: FQG61/FQG62.Details
52013061O-ringSource holder QG020/QG100 chemical versionDetails
XPRFID-Spare RFID TAGConsists of: RFID TAG, fixing wire, crimp sleeve.Enquire

Configurator / Comparison

ConfigureEnter a product model to configure
CompareEnter the product models to compare — up to three products

Source: Endress+Hauserpublic information from the website

All documents for this model
4 documents · click to expand
BA

BA 源盒FQG61、FQG62 Chinese

Rev. 0125 92 pages
Preview
BA

BA Source container FQG61, FQG62 English

Rev. 0125 100 pages
Preview
TI

源盒FQG61、FQG62 技术资料 Chinese

Rev. 1919 56 pages
Preview
TI

Source container FQG61, FQG62 Technical Information English

Rev. 1919 56 pages
Preview
Municipal Wastewater Treatment

Municipal Wastewater Treatment · radiant — Application Case

Site Conditions

This is a Municipal Wastewater Treatment customer whose plant is at the inlet, biological and outlet stages of a municipal plant. The main task was non-contact level measurement in extreme duty using radiometric principles.

The Problem

The difficulty lies in high solids and corrosive fractions, with large diurnal flow swings; at the same time effluent parameters drift and probes foul with sludge and biofilm. Specification therefore means sizing source strength against wall thickness, density and licensing requirements.

What We Did

We took a level range approach — level solution matched to the actual tank and silo geometry. Delivery was built around sizing source strength against wall thickness, density and licensing requirements, with point-by-point checks on: stable compliance with Grade 1A discharge limits and online data reporting.

Operating Outcome

In service the system performs as intended — measurable under high temperature, pressure and corrosion without intruding into the vessel, with routine maintenance falling noticeably.

Food & Beverage

Food & Beverage · radiant — Application Case

Application Context

dosing, pasteurising, filling and CIP stages is a typical Food & Beverage duty. The project started from radiant, focusing on non-contact level measurement in extreme duty using radiometric principles.

Process Constraints

frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles, and syrups, dairy and beverages that scale easily, under strict hygienic requirements only amplify it. The radiant specification must therefore cover sizing source strength against wall thickness, density and licensing requirements.

Delivery Approach

Delivery was handled as the level range — level solution matched to the actual tank and silo geometry. Execution centred on sizing source strength against wall thickness, density and licensing requirements, together with food-safety and hygienic design per 3-A / EHEDG.

Field Results

Operation shows the benefit clearly — measurable under high temperature, pressure and corrosion without intruding into the vessel, with confidence in this measuring point improving markedly.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · radiant — Application Case

Customer Profile

The typical setting in Chemical & Pharmaceutical is reactors, metering vessels and finished-product tanks. The measuring point is radiant; the focus is non-contact level measurement in extreme duty using radiometric principles.

Key Difficulties

Conditions are highly corrosive media with solvents and crystallisation, partly in explosive atmospheres and, on top of that, corrosion plus Ex protection shortens instrument life and makes downtime costly, which a standard configuration cannot handle for long. The design therefore starts with sizing source strength against wall thickness, density and licensing requirements.

Implementation

The scope ran through the level range: level solution matched to the actual tank and silo geometry. Technically it hinges on sizing source strength against wall thickness, density and licensing requirements, and on these requirements: Ex certification and GMP traceability of records.

Customer Benefit

In practice this means measurable under high temperature, pressure and corrosion without intruding into the vessel, which cut the workload on the site maintenance team.

Power & Energy

Power & Energy · radiant — Application Case

Project Background

The project covers boiler feedwater, turbine condensate and FGD/DeNOx systems. The work centres on radiant and the core task is non-contact level measurement in extreme duty using radiometric principles.

Operating Challenges

The medium and duty are characterised by high temperature and pressure with both steam and saline water, and drift under high temperature and pressure, with very costly unplanned outages. So radiant cannot be specified generically — it calls for sizing source strength against wall thickness, density and licensing requirements.

Solution

The Ainstru engineering team worked through the level range: level solution matched to the actual tank and silo geometry. The configuration involves sizing source strength against wall thickness, density and licensing requirements, with these requirements in view: compliance with power-industry codes and unit availability targets.

Results

Since start-up the system has run steadily; the practical result is measurable under high temperature, pressure and corrosion without intruding into the vessel.