Technical Data
CompareConductivity
| Measuring Principle | Inductive |
| Application | - Alkali/acid concentration control - phase separation in pipe systems - monitoring and control of bottle cleaning systems - product monitoring in breweries, dairies and the beverage industry - CIP system control |
| Characteristic | Compact measuring system for an inductive conductivity measurement special made for media with a high conductivity. |
| Measuring range | 100 µS/cm - 2000 mS/cm |
| Measuring principle | Inductive measuring system including transmitter and sensor. |
| Design | Compact transmitter made of stainless steel with an inductive sensor made of PEEK in a separate or compact version. |
| Material | Housing: Stainless steel 1.4301 polished Window: Polycarbonate Sensor: PEEK, stainless steel 1.4435 Sealing: Chemraz |
| Process temperature | max. 140°C (max. 30 min.) 284°F (max. 30 min.) |
| Process pressure | max. 12 bar (90°C) (174 psi(194°F)) |
| Temperature sensor | Pt100 |
| Ex certification | No |
| Ingress protection | IP67 |
| Input | One-channel transmitter incl. temperature. |
| Output / communication | Conductivity and temperature 0/4-20 mA, HART, Profibus. |
Field of application
Smartec CLD132 is a measuring system for inductive conductivity. The combination of a transmitter and a sensor is interference-free, easy-to-use and features a stainless steel temperature sensor that gives you an ultrafast temperature response for reliable measurement in separation applications.
Smartec CLD132 is suited to applications such as:
Alkaline/acid concentration control in remaking
Phase separation of product/water and product/product mixtures in pipe systems
Product monitoring
Smartec CLD132 features the following protocols and interfaces:
0/4...20mA
HART
PROFIBUS DP
PROFIBUS PA
Benefits
Engineered for quick temperature response
Low risk of soiling due to encapsulated sensor
Insensitive to polarization
Conductivity:
| Measuring Principle | Inductive | |
| Application | - Alkali/acid concentration control - phase separation in pipe systems - monitoring and control of bottle cleaning systems - product monitoring in breweries, dairies and the beverage industry - CIP system control | |
| Characteristic | Compact measuring system for an inductive conductivity measurement special made for media with a high conductivity. | |
| Measuring range | 100 µS/cm - 2000 mS/cm | |
| Measuring principle | Inductive measuring system including transmitter and sensor. | |
| Design | Compact transmitter made of stainless steel with an inductive sensor made of PEEK in a separate or compact version. | |
| Material | Housing: Stainless steel 1.4301 polished Window: Polycarbonate Sensor: PEEK, stainless steel 1.4435 Sealing: Chemraz | |
| Process temperature | max. 140°C (max. 30 min.) 284°F (max. 30 min.) |
|
| Process pressure | max. 12 bar (90°C) (174 psi(194°F)) |
|
| Temperature sensor | Pt100 | |
| Ex certification | No | |
| Ingress protection | IP67 | |
| Input | One-channel transmitter incl. temperature. | |
| Output / communication | Conductivity and temperature 0/4-20 mA, HART, Profibus. | |
Spare Parts & Accessories 46
46 spare parts · click to collapse
| Order code | Description | |
|---|---|---|
| 51501574 | CLD132/134 Split-Type Housing BaseKit CLD132/4 lower casing seperate vers. | Details |
| 51501576 | CLD132/134 Compact Housing BaseKit CLD132/4 lower casing compact vers. | Details |
| 51501577 | CLD132/134 Integrated Housing PanelKit CLD132/4 housing cover complete | Details |
| 50062121 | Transmitter protective coverround post installation | Details |
| 51501589 | CLD132/4 PG cable entry and breather holeKit CLD132/4 PG glands+Goretex filter | Details |
| 51502282 | CLD132/4 M20 cable entry and breather holeKit CLD132/4 M20 glands+Goretex filter | Details |
| 51502283 | CLD132/4 coil cable entry and breather holeKit CLD132/4 Contuit glands+Goretex | Details |
| 51501596 | CLD132/4 Seals and small partsKit CLD132/4 Sealings and small parts | Details |
| 51502382 | CLD132/4 Relay Wiring Protective CoverKit CLD132/4 Protection against contact | Details |
| 51501578 | Kit CLD132 Integrated Type APVKit CLD132 sensor compact version APV | Details |
| 51501579 | Kit CLD132 Compact Type Clamp 2"Kit CLD132 sensor compact ver. Clamp 2" | Details |
| 51501580 | Kit CLD132 Integral Type G1.5"Kit CLD132 sensor compact version G1.5" | Details |
| 51501581 | Kit CLD132 integrated hygienic connection DN50Kit CLD132 sensor compact dairy fit.DN50 | Details |
| 51501582 | Kit CLD132 Integrated VariventKit CLD132 sensor compact vers. Varivent | Details |
| 51501583 | Kit CLD132 sensor compact ver. Perlic 2"Kit CLD132 sensor compact ver. Perlic 2" | Details |
| 51502279 | Kit CLD132 Integrated SMSKit CLD132 sensor compact version SMS | Details |
| 71020489 | sensor | Details |
| 71020487 | sensorKit CLD134, version MV5, spare sensor | Details |
| 71020490 | sensor | Details |
| 71020491 | sensor | Details |
Configurator / Comparison
Manuals & Documents 8
All documents for this model8 documents · click to expand
BA Smartec CLD132/134 Chinese
BA Smartec CLD132 Chinese
BA Smartec CLD132/134 English
BA Smartec CLD132 English
Operating Instructions SmarTec S CLD 132 Hart English
Operating Instructions Smartec S CLD132 English
Operating Instructions Smartec S CLD132 Profibus- PA/-DP English
Technical Information Smartec S CLD132 English
Case Studies
Municipal Wastewater Treatment · conductivity — Application Case
Customer Profile
The typical setting in Municipal Wastewater Treatment is the inlet, biological and outlet stages of a municipal plant. The measuring point is conductivity; the focus is measuring conductivity to judge water purity or chemical concentration.
Key Difficulties
Conditions are high solids and corrosive fractions, with large diurnal flow swings and, on top of that, effluent parameters drift and probes foul with sludge and biofilm, which a standard configuration cannot handle for long. The design therefore starts with choosing two- or four-electrode cells by range and corrosivity.
Implementation
The scope ran through the analyser range: selecting as part of a complete water-analysis system. Technically it hinges on choosing two- or four-electrode cells by range and corrosivity, and on these requirements: stable compliance with Grade 1A discharge limits and online data reporting.
Customer Benefit
In practice this means stable concentration control and less reagent and discharge waste, which cut the workload on the site maintenance team.
Food & Beverage · conductivity — Application Case
Project Background
The project covers dosing, pasteurising, filling and CIP stages. The work centres on conductivity and the core task is measuring conductivity to judge water purity or chemical concentration.
Operating Challenges
The medium and duty are characterised by syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. So conductivity cannot be specified generically — it calls for choosing two- or four-electrode cells by range and corrosivity.
Solution
The Ainstru engineering team worked through the analyser range: selecting as part of a complete water-analysis system. The configuration involves choosing two- or four-electrode cells by range and corrosivity, with these requirements in view: food-safety and hygienic design per 3-A / EHEDG.
Results
Since start-up the system has run steadily; the practical result is stable concentration control and less reagent and discharge waste.
Chemical & Pharmaceutical · conductivity — Application Case
Project Overview
The site is reactors, metering vessels and finished-product tanks. The customer wanted conductivity properly configured on the existing plant, with the aim of measuring conductivity to judge water purity or chemical concentration.
Technical Challenges
What actually causes trouble is highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. All of this means the conductivity configuration has to involve choosing two- or four-electrode cells by range and corrosivity.
Configuration
Supply and installation were organised through the analyser range: selecting as part of a complete water-analysis system. On site this meant choosing two- or four-electrode cells by range and corrosivity, checked line by line against: Ex certification and GMP traceability of records.
Performance
After commissioning the outcome is stable concentration control and less reagent and discharge waste, and the readings are now used for process tuning.
Power & Energy · conductivity — Application Case
Site Conditions
This is a Power & Energy customer whose plant is at boiler feedwater, turbine condensate and FGD/DeNOx systems. The main task was measuring conductivity to judge water purity or chemical concentration.
The Problem
The difficulty lies in high temperature and pressure with both steam and saline water; at the same time drift under high temperature and pressure, with very costly unplanned outages. Specification therefore means choosing two- or four-electrode cells by range and corrosivity.
What We Did
We took a analyser range approach — selecting as part of a complete water-analysis system. Delivery was built around choosing two- or four-electrode cells by range and corrosivity, with point-by-point checks on: compliance with power-industry codes and unit availability targets.
Operating Outcome
In service the system performs as intended — stable concentration control and less reagent and discharge waste, with routine maintenance falling noticeably.