Technical Data
CompareConductivity
| Measuring Principle | |
| Application | Pure and ultrapure water, pharmaceutical industry |
| Characteristic | Hygienic 2-electrode conductivity sensor |
| Measuring range | k=0,1: 0,04 to 500 µS/cm |
| Measuring principle | Conductivity cell |
| Design | Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished stainless steel and FDA sealing material; sterilizable |
| Material | Isolation: PEEK Electrode: polished stainless steel 1.4435 Sealing: Isolast (FFKM) FDA conform |
| Dimension | Electrode diameter: 17 mm (0.66 inch) Electrode length: 54 mm (2.13 inch) |
| Process temperature | -5 to 120 °C (23 to 248 °F) For Sterilizsation: max. 150 °C at 5 bar (Max. 302 °F at 73 psi) |
| Process pressure | 13 bar at 20 °C (188 psi at 68 °F) 9 bar at 120 °C (130 psi at 248 °F) |
| Temperature sensor | NTC 30K |
| Connection | Process connection: Varivent, Neumo, Clamp Cable: Memosens connector |
| Ingress protection | IP68 |
| Certificates, declarations and approvals | Material certification 3.1 EHEDG certified, hygienic design, CoC ASME BPE Conformity to cGMP derived requirements ASME BPE FDA approved Regulation (EC) No. 1935/2004 China Food |
The Condumax CLS16D is a high-end digital conductivity sensor with hygienic design approval. For optimal measurement processes and product quality, it provides high-precision measurement and reliable data. Numerous approvals and various hygienic connections ensure a perfect fit for the measuring process. With Memosens digital technology, the CLS16D combines maximum process and data integrity with simple operation. It is corrosion- and moisture-resistant and can be calibrated in the laboratory for sensor pre-maintenance.
Benefits
-
Highest measuring accuracy
-
Hygienic design, EHEDG-certified
-
Wetted parts materials are biocompatibility-certified in accordance with USP Cl. VI and FDA standards
-
Can be sterilized and disinfected, and withstands autoclaving
-
Quality certificate with cell constant documentation
-
Stores sensor-specific parameters for traceability and predictive maintenance
-
Convenient maintenance strategy with simple calibration and sensor management using Memobase Plus
Application areas
The Condumax CLS16D measures the conductivity of pure and ultrapure water for monitoring and control:
-
Ion exchanger
-
Reverse osmosis
-
Distillation
-
Electrodeionization
-
Water for injection (WFI)
-
With various international explosion-protection certifications such as ATEX, FM, CSA and NEPSI, the sensor is suitable for hazardous areas
Spare Parts & Accessories 2
2 spare parts · click to collapse
| Order code | Description | |
|---|---|---|
| 50081902 | Calibration solution 74.0 µS/cm500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO | Details |
| 50081903 | calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO. | Details |
Configurator / Comparison
A
Manuals & Documents 6
All documents for this model6 documents · click to expand
Case Studies
Municipal Wastewater Treatment · conductivity — Application Case
Customer Profile
The typical setting in Municipal Wastewater Treatment is the inlet, biological and outlet stages of a municipal plant. The measuring point is conductivity; the focus is measuring conductivity to judge water purity or chemical concentration.
Key Difficulties
Conditions are high solids and corrosive fractions, with large diurnal flow swings and, on top of that, effluent parameters drift and probes foul with sludge and biofilm, which a standard configuration cannot handle for long. The design therefore starts with choosing two- or four-electrode cells by range and corrosivity.
Implementation
The scope ran through the analyser range: selecting as part of a complete water-analysis system. Technically it hinges on choosing two- or four-electrode cells by range and corrosivity, and on these requirements: stable compliance with Grade 1A discharge limits and online data reporting.
Customer Benefit
In practice this means stable concentration control and less reagent and discharge waste, which cut the workload on the site maintenance team.
Food & Beverage · conductivity — Application Case
Project Background
The project covers dosing, pasteurising, filling and CIP stages. The work centres on conductivity and the core task is measuring conductivity to judge water purity or chemical concentration.
Operating Challenges
The medium and duty are characterised by syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. So conductivity cannot be specified generically — it calls for choosing two- or four-electrode cells by range and corrosivity.
Solution
The Ainstru engineering team worked through the analyser range: selecting as part of a complete water-analysis system. The configuration involves choosing two- or four-electrode cells by range and corrosivity, with these requirements in view: food-safety and hygienic design per 3-A / EHEDG.
Results
Since start-up the system has run steadily; the practical result is stable concentration control and less reagent and discharge waste.
Chemical & Pharmaceutical · conductivity — Application Case
Project Overview
The site is reactors, metering vessels and finished-product tanks. The customer wanted conductivity properly configured on the existing plant, with the aim of measuring conductivity to judge water purity or chemical concentration.
Technical Challenges
What actually causes trouble is highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. All of this means the conductivity configuration has to involve choosing two- or four-electrode cells by range and corrosivity.
Configuration
Supply and installation were organised through the analyser range: selecting as part of a complete water-analysis system. On site this meant choosing two- or four-electrode cells by range and corrosivity, checked line by line against: Ex certification and GMP traceability of records.
Performance
After commissioning the outcome is stable concentration control and less reagent and discharge waste, and the readings are now used for process tuning.
Power & Energy · conductivity — Application Case
Site Conditions
This is a Power & Energy customer whose plant is at boiler feedwater, turbine condensate and FGD/DeNOx systems. The main task was measuring conductivity to judge water purity or chemical concentration.
The Problem
The difficulty lies in high temperature and pressure with both steam and saline water; at the same time drift under high temperature and pressure, with very costly unplanned outages. Specification therefore means choosing two- or four-electrode cells by range and corrosivity.
What We Did
We took a analyser range approach — selecting as part of a complete water-analysis system. Delivery was built around choosing two- or four-electrode cells by range and corrosivity, with point-by-point checks on: compliance with power-industry codes and unit availability targets.
Operating Outcome
In service the system performs as intended — stable concentration control and less reagent and discharge waste, with routine maintenance falling noticeably.