Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

CLS52 Indumax

CLS52 Indumax
Indumax CLS52
Configurable Models 12 models
The Indumax CLS52 inductive conductivity sensor is equipped with a stainless steel temperature sensor that gives you an ultrafast temperature response to measure exactly in separation applications.
Compare

Conductivity

Measuring Principle Inductive
Application Food, pharmaceuticals
Characteristic Hygienic inductive measuring cell for conductivity measurements, concentration control, CIP-control, cleaning and phase separation.
Measuring range 0.1 to 2000 mS/cm
Measuring principle Inductive conductivity measurement.
Design Hygienic inductive conductivity cell with an outside temperature sensor for a faster temperature reading.
Material Sensor: PEEK
Heat contact nut: stainless steel 1.4435
Sealing: Chemraz
Dimension Coil: outside diameter: appr. 47 mm inside diameter: appr. 15 mm
Process temperature Max. 125°C (short time up to 140°C)
max. 257°F (short time up to 284°F)
Process pressure Max. 16 bar at 90 °C
(Max. 232 psi at 194 °F)
Temperature sensor Integrated Pt100 in a heat contact nut
Ex certification not available as Ex version
Connection Process connections: dairy fitting, Clamp, thread 1,5", Varivent, APV, Perlick, SMS cable: fixed cable connection.
Ingress protection IP67
Certificates, declarations and approvals Calibration certification.

The CLS52 inductive conductivity sensor is used in the food and beverage industry to protect products and production processes. In separated measurement applications, a stainless steel temperature sensor provides ultra-fast temperature response for precise measurement.

 

Advantages

Fast temperature response

Hygienic design prevents contamination

Suitable for CIP and SIP cleaning

Encapsulated sensor construction prevents contamination

Insensitive to polarization

 

Areas of application

The Indumax CLS52 is used for toroidal conductivity measurement, especially in food&beverage applications requiring fast temperature response, e.g.:

 

Acid/alkali concentration control during preparation

Separation between medium/water and medium/medium mixtures in pipes

Monitoring and control of bottle washing systems

Product monitoring in breweries, dairies and the beverage industry

CIP system control

Spare Parts & Accessories 6

6 spare parts · click to collapse
Order codeDescription
51500753Cable accessories CLS50/52Details
50081902Calibration solution 74.0 µS/cm500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISODetails
50081903calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO.Details
50081904Calibration solution 1.406 mS/cm500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISODetails
50081905calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM to NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISODetails
50081906calibration solution500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISODetails

Configurator / Comparison

ConfigureEnter a product model to configure
CompareEnter the product models to compare — up to three products

   

   

 

 

A

Source:Endress+Hauserwebsite public information

Measuring principle video

Click to play
12 models · click to collapse
All documents for this model
6 documents · click to expand
BA

Indumax CLS52 操作手册 Chinese

Rev. 0217 24 pages
Preview
BA

Indumax CLS52 Operating Instructions English

Rev. 0217 24 pages
Preview
TI

技术资料 Indumax CLS52 Chinese

Rev. 1316 16 pages
Preview
TI

技术资料 Indumax H CLS52 Chinese

Rev. 0704 8 pages
Preview
TI

Indumax CLS52 Technical Information English

Rev. 1316 16 pages
Preview
TI

Technical Information CLS52 English

12 pages
Preview
Municipal Wastewater Treatment

Municipal Wastewater Treatment · conductivity — Application Case

Customer Profile

The typical setting in Municipal Wastewater Treatment is the inlet, biological and outlet stages of a municipal plant. The measuring point is conductivity; the focus is measuring conductivity to judge water purity or chemical concentration.

Key Difficulties

Conditions are high solids and corrosive fractions, with large diurnal flow swings and, on top of that, effluent parameters drift and probes foul with sludge and biofilm, which a standard configuration cannot handle for long. The design therefore starts with choosing two- or four-electrode cells by range and corrosivity.

Implementation

The scope ran through the analyser range: selecting as part of a complete water-analysis system. Technically it hinges on choosing two- or four-electrode cells by range and corrosivity, and on these requirements: stable compliance with Grade 1A discharge limits and online data reporting.

Customer Benefit

In practice this means stable concentration control and less reagent and discharge waste, which cut the workload on the site maintenance team.

Food & Beverage

Food & Beverage · conductivity — Application Case

Project Background

The project covers dosing, pasteurising, filling and CIP stages. The work centres on conductivity and the core task is measuring conductivity to judge water purity or chemical concentration.

Operating Challenges

The medium and duty are characterised by syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. So conductivity cannot be specified generically — it calls for choosing two- or four-electrode cells by range and corrosivity.

Solution

The Ainstru engineering team worked through the analyser range: selecting as part of a complete water-analysis system. The configuration involves choosing two- or four-electrode cells by range and corrosivity, with these requirements in view: food-safety and hygienic design per 3-A / EHEDG.

Results

Since start-up the system has run steadily; the practical result is stable concentration control and less reagent and discharge waste.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · conductivity — Application Case

Project Overview

The site is reactors, metering vessels and finished-product tanks. The customer wanted conductivity properly configured on the existing plant, with the aim of measuring conductivity to judge water purity or chemical concentration.

Technical Challenges

What actually causes trouble is highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. All of this means the conductivity configuration has to involve choosing two- or four-electrode cells by range and corrosivity.

Configuration

Supply and installation were organised through the analyser range: selecting as part of a complete water-analysis system. On site this meant choosing two- or four-electrode cells by range and corrosivity, checked line by line against: Ex certification and GMP traceability of records.

Performance

After commissioning the outcome is stable concentration control and less reagent and discharge waste, and the readings are now used for process tuning.

Power & Energy

Power & Energy · conductivity — Application Case

Site Conditions

This is a Power & Energy customer whose plant is at boiler feedwater, turbine condensate and FGD/DeNOx systems. The main task was measuring conductivity to judge water purity or chemical concentration.

The Problem

The difficulty lies in high temperature and pressure with both steam and saline water; at the same time drift under high temperature and pressure, with very costly unplanned outages. Specification therefore means choosing two- or four-electrode cells by range and corrosivity.

What We Did

We took a analyser range approach — selecting as part of a complete water-analysis system. Delivery was built around choosing two- or four-electrode cells by range and corrosivity, with point-by-point checks on: compliance with power-industry codes and unit availability targets.

Operating Outcome

In service the system performs as intended — stable concentration control and less reagent and discharge waste, with routine maintenance falling noticeably.