Instrumentation · Engineering · Package · PLC
0755-8222 1901 sales@ainstru.com

PMD55 differential pressure transmitter

PMD55 differential pressure transmitter

Deltabar M PMD55

Configurable Models 19 models
PMD55 is a commonly used E+H differential pressure transmitter, suitable for most application scenarios and cost-effective.
Compare

Continuous / Liquids

Measuring Principle Differential pressure
Characteristic / Application Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
Supply / Communication 4...20mA HART:
11,5...45V DC
Ex ia: 11,5...30V DC
Accuracy 0,1%
"PLATINUM" 0,075%
Long term stability 0,05% of URL/year
0,125% of URL/5 years
Ambient temperature -40°C...85°C
(-40°F...185°F)
Process temperature -40 °C...85 °C
(-40 °F...185 °F)
Process pressure / max. overpressure limit 160 bar
Pressure measuring range 10mbar...40bar
(0.15...580psi)
Main wetted parts 316L
Process connection 1/4-18 NPT
Communication 4...20mA HART
PROFIBUS PA
FOUNDATION Fieldbus
Certificates / Approvals ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI
Safety approvals SIL
Design approvals EN 10204-3.1
NACE MR0175, MR0103
AD2000
Options 4-line digital display
Aluminium housing
Successor PMD55B
Application limits Measuring cell:
Metal welded

Gas

Measuring Principle Differential pressure
Product headline Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
Max. measurement error 0,1%
"PLATINUM" 0,075%
Max. process pressure 10mbar...40bar
(0.15...580psi)
Medium temperature range Temperature gradient from pressure piping
Display/Operation Option
Outputs 4...20mA HART
Digital communication HART
Hazardous area approvals ATEX, FM, CSA, IECEx, INMETRO, NEPSI, TIIS
Functional safety SIL
Material certificates NACE MR0103
NACE MR0175
EN10204-3.1

Liquids

Measuring Principle Differential pressure
Product headline Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
Max. measurement error 0,1%
"PLATINUM" 0,075%
Max. process pressure 10 mbar...40 bar
(0.15...580 psi)
Medium temperature range Temperature gradient from pressure piping
Display/Operation Option
Outputs 4...20mA HART
Digital communication HART
Hazardous area approvals ATEX, FM, CSA, IECEx, INMETRO, NEPSI, TIIS
Functional safety SIL
Material certificates NACE MR0103
NACE MR0175
EN10204-3.1

Pressure

Measuring Principle Differential pressure
Characteristic Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
Supply voltage 4...20 mA HART
11,5...45V DC (Non Ex):
Ex ia: 11,5...30V DC
PROFIBUS PA:
9...32 V DC (Non Ex)
FOUNDATION Fieldbus:
9...32 V DC (Non Ex)
Reference Accuracy Standard 0.1%
Platinum 0.075%
Long term stability 0.05% of URL/ year
0.13% of URL/ 5 years
0.23% of URL/ 10 years
Process temperature -40°C...+85°C
(-40°F...+185°F)
Ambient temperature -40°C...+85°C
(-40°F...+185°F)
Measuring cell 10 mbar...40 bar
(0.15...580 psi)
Smallest calibratable span 10 mbar (0.15 psi)
Max. Turn down 20:1
Max. overpressure limit on one side:
160 bar
(2300 psi)
Process connection 1/4-18 NPT
Material process membrane 316L, AlloyC,
Material gasket Viton, PTFE, EPDM, NBR
Fill fluid Silicone oil
Inert oil
Material housing Die-cast aluminum
Communication 4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
Certificates / Approvals ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, UK Ex
Safety approvals SIL
Design approvals NACE MR0175
EN10204-3.1
Successor PMD55B

Steam

Measuring Principle Differential pressure
Product headline Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
Max. measurement error 0,1%
"PLATINUM" 0,075%
Max. process pressure 10mbar...40bar
(0.15...580psi)
Medium temperature range Temperature gradient from pressure piping
Display/Operation Option
Outputs 4...20mA HART
Digital communication HART
Hazardous area approvals ATEX, FM, CSA, IECEx, INMETRO, NEPSI, TIIS
Functional safety SIL
Material certificates NACE MR0103
NACE MR0175
EN10204-3.1

The Deltabar PMD55 differential pressure transmitter uses a piezoresistive sensor and a welded metal diaphragm and is mainly used for differential pressure measurement of liquids, steam and gases in the process industry or the environmental protection industry. The measuring range is adjustable, and fast setup simplifies commissioning, reduces costs and saves time. It complies with IEC61508/IEC61511 and meets SIL3 safety requirements.

Advantages

Reference accuracy: 0.1%, Platinum version: up to ±0.075%

Turndown ratio: 100:1 or higher

One-piece transmitter

Fast commissioning using DIP switches

Simple, safe menu-guided operation: - via the on-site display module: - via 4...20 mA HART - via PROFIBUS PA, - via FF FOUNDATION Fieldbus

Modular continuous measurement of differential pressure, static pressure and pressure (Deltabar M - Deltapilot M, Cerabar M), e.g. - replaceable display, - universal electronics for pressure and differential pressure measurement

Wide range of approvals for international use

Application areas

For liquid level, volume or mass measurement, differential pressure monitoring of filters and pumps, and, in combination with primary elements, measurement of gas, steam and liquid flow (volume flow or mass flow)

Process connection: thread

Process temperature: -40... 85°C (-40... 185°F)

Measuring range: 0.5mbar ... 40bar (0.0072 ... 600psi)

Accuracy: ±0.1%, "Platinum" ±0.075% (optional)

International explosion protection approval, SIL

 

 

E+H Product Model Configurator

Please enter the product model to configure

Product comparison

Please enter the product models to compare in the input fields below; up to three products can be compared

 

   

Technical documentation

   

Technical data

 

 

Documents / Manuals / Software

 

 

Accessories / Spare parts

 

   

Orifice plate flowmeter measuring principle video

   

Pitot tube measuring principle video

Source:Endress+Hauserpublic information from the website

Model Name
PMD55-AA21BA67HGCHAJA1A+AK 差压变送器
PMD55-AA21BA27CGCHA4A1A+AK 差压变送器
PMD55-AA21BA67DGCHAJA1A+AK 差压变送器
PMD55-AA21BA67DGJHAJA1AABPDZ1 差压变送器
PMD55-AA21BB67DGJHA4A1APBZ1 差压变送器
PMD55-AA21BA67LGCHAJA1A+AK 差压变送器
PMD55-NA21BA67LGCHAJA1A+AK 差压变送器
PMD55-BA21BA67LGCHAJA1A+AK 差压变送器
PMD55-BA21BA27CGCHAJA1A+AK 差压变送器
PMD55-AA22AA67FGJHAJA1J 差压变送器
PMD55-BC21BB67LGCHAJA1A+AK 差压变送器
PMD55-BA21BA67GGCHAJA1A+AK 差压变送器
PMD55-BA21BA67HGCHAJA1A+AK 差压变送器
PMD55-NA21BA67GGCHAJA1A+AK 差压变送器
PMD55-NA21BB67HGCHAJA1A+AK 差压变送器
PMD55-NA21BA67HGCHAJA1A+AK 差压变送器
PMD55-BD22AB77LGJHAJA1J+P5 差压变送器
PMD55-AA21BA67DGJHAJA1A+AK 差压变送器
PMD55-AA21BA67DGCHAJA2A+AA 差压变送器
Municipal Wastewater Treatment

Municipal Wastewater Treatment — Wastewater Treatment Plant · Differential Pressure Measurement Application Case

Project Background

Municipal Wastewater Treatment customers at the Wastewater Treatment Plant stage have long faced the challenge of insufficient accuracy in effluent quality monitoring and instruments prone to contamination causing data drift, affecting compliance discharge, placing higher demands on the stability and measurement accuracy of Differential Pressure Measurement.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Differential Pressure Measurement for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Municipal Wastewater Treatment conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Food & Beverage

Food & Beverage — Food Production Line · Differential Pressure Measurement Application Case

Project Background

Food & Beverage customers at the Food Production Line stage have long faced the challenge of high hygienic measurement requirements, with inline instruments needing to withstand CIP/SIP cleaning while maintaining stable accuracy, placing higher demands on the stability and measurement accuracy of Differential Pressure Measurement.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Differential Pressure Measurement for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Food & Beverage conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Chemical & Pharmaceutical

Chemical & Pharmaceutical — Reactors and Storage Tanks · Differential Pressure Measurement Application Case

Project Background

Chemical & Pharmaceutical customers at the Reactors and Storage Tanks stage have long faced the challenge of highly corrosive media and complex conditions, requiring corrosion-resistant and explosion-proof instruments, placing higher demands on the stability and measurement accuracy of Differential Pressure Measurement.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Differential Pressure Measurement for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Chemical & Pharmaceutical conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Power & Energy

Power & Energy — Power Plant Boiler and Turbine · Differential Pressure Measurement Application Case

Project Background

Power & Energy customers at the Power Plant Boiler and Turbine stage have long faced the challenge of unstable measurement under high temperature and pressure, with high maintenance and downtime costs, placing higher demands on the stability and measurement accuracy of Differential Pressure Measurement.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Differential Pressure Measurement for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Power & Energy conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.