Measuring method
-
1. The water sample enters the reactor and reacts with acid, at which point the inorganic carbon in the water sample is converted into carbon dioxide;
-
2. Using carbon dioxide-free air (pure oxygen) as the carrier gas, the carbon dioxide converted from inorganic carbon is purged out, and at this point the total inorganic carbon (TIC) can be measured by detection with an NDIR detector;
-
3. The water sample from which TIC (total inorganic carbon) has been removed is then oxidized;
-
4. After oxidation is complete, the carbon dioxide converted from organic carbon is purged out by the carrier gas and detected by the NDIR detector to obtain the total organic carbon (TOC) content.
Instrument features
-
Adopts an oxidation method combining UV persulfate and hydroxyl radicals, resulting in higher oxidation efficiency and high accuracy even at low concentrations
-
Adopts a sample inlet line with a 4 mm inner diameter, allowing passage of soft particles up to 2 mm in diameter, and comes with a built-in flow cell, so routine samples require no filtration
-
The two-point automatic calibration function reduces additional maintenance workload for operators, and sequential batch measurement reduces reagent consumption.
-
The analysis process does not use alkaline reagents, effectively reducing the difficulty for operators to prepare reagents and eliminating measurement interference caused by carbon dioxide dissolving into alkaline reagents
-
Low maintenance workload, low maintenance cost
-
Friendly human-machine interface: equipped with a high-definition touchscreen, featuring a user-friendly operating interface and instrument interfaces, enabling quick and convenient data download and software updates, and a multi-level management mode for convenient instrument management by users
Technical specifications
-
Measuring parameters TOC (NPOC)/TIC
-
Oxidation method Oxidation method combining UV persulfate and hydroxyl radicals
-
Measuring method After the water sample is oxidized, the carbon dioxide concentration is measured with an NDIR detector
-
Measuring range 0-10 mg/L
-
Minimum detection limit 0.05 mg/L
-
Repeatability ±2% of reading or ±0.05 mg/L (whichever is greater)
-
Accuracy ±3% of reading or ±0.05 mg/L (whichever is greater)
-
Zero drift Within ±2%f.s./day
-
Span drift Within ±2%f.s./day
-
Analysis time 7 minutes (TOC)
-
Measuring interval Adjustable (in minutes)
-
Interference Possible turbidity, fat, oil, protein, and tar may interfere with the measurement results
-
Sample inlet requirements Flow rate: ≥100 mL/min
-
Particle size: ≤2 mm, soft particles
-
Sample inlet pressure: ≤1 bar
-
Carrier gas (oxygen) Oxygen generator: power supply voltage and frequency: 220V AC/50Hz (±10%)
-
Oxygen concentration: >92%
-
Pressure: >0.05 Mpa
-
Oxygen consumption flow rate: maximum 1.7L/min
-
Water sample temperature 5℃– 45℃
-
Maintenance interval Once every 3 months
-
Protection class IP44
-
Signal output 0/4~20 mA, Modbus/RTU RS485
-
Relay output 4 relays (24v 2A, dry contact)
-
Display 7-inch, 800 × 480 pixels (4:3) TFT-LCD touchscreen
-
Data storage 9999 analysis data records, 99 fault messages
-
Power supply 100 – 240 V AC, 50/60Hz
-
Dimensions 1100 x 560 x 370 mm
-
Weight 65 Kg
Ordering information
Main unit
8653100 QL3580M online total organic carbon analyzer
Accessories
-
8663500 Oxygen generator
-
Other spare parts and annual maintenance kit
-
8651700 QL3580M annual maintenance kit
-
8651000 NDIR carbon dioxide detector 343
-
9789600 Circulation pump dedicated tubing kit
-
8654000 Peristaltic pump tubing kit
Source: Public information from the HACH water quality website
Case Studies
Municipal Wastewater Treatment · HACH-ONLINE — Application Case
Project Background
The project covers the inlet, biological and outlet stages of a municipal plant. The work centres on HACH-ONLINE and the core task is continuous outfall monitoring with Hach online instruments.
Operating Challenges
The medium and duty are characterised by high solids and corrosive fractions, with large diurnal flow swings, and effluent parameters drift and probes foul with sludge and biofilm. So HACH-ONLINE cannot be specified generically — it calls for configured by monitoring range, reagent replacement interval and conditioning.
Solution
The Ainstru engineering team worked through the online water-analysis line: integrating around each online monitoring point. The configuration involves configured by monitoring range, reagent replacement interval and conditioning, with these requirements in view: stable compliance with Grade 1A discharge limits and online data reporting.
Results
Since start-up the system has run steadily; the practical result is steady reagent and consumable supply, making upkeep predictable.
Food & Beverage · HACH-ONLINE — Application Case
Project Overview
The site is dosing, pasteurising, filling and CIP stages. The customer wanted HACH-ONLINE properly configured on the existing plant, with the aim of continuous outfall monitoring with Hach online instruments.
Technical Challenges
What actually causes trouble is syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. All of this means the HACH-ONLINE configuration has to involve configured by monitoring range, reagent replacement interval and conditioning.
Configuration
Supply and installation were organised through the online water-analysis line: integrating around each online monitoring point. On site this meant configured by monitoring range, reagent replacement interval and conditioning, checked line by line against: food-safety and hygienic design per 3-A / EHEDG.
Performance
After commissioning the outcome is steady reagent and consumable supply, making upkeep predictable, and the readings are now used for process tuning.
Chemical & Pharmaceutical · HACH-ONLINE — Application Case
Site Conditions
This is a Chemical & Pharmaceutical customer whose plant is at reactors, metering vessels and finished-product tanks. The main task was continuous outfall monitoring with Hach online instruments.
The Problem
The difficulty lies in highly corrosive media with solvents and crystallisation, partly in explosive atmospheres; at the same time corrosion plus Ex protection shortens instrument life and makes downtime costly. Specification therefore means configured by monitoring range, reagent replacement interval and conditioning.
What We Did
We took a online water-analysis line approach — integrating around each online monitoring point. Delivery was built around configured by monitoring range, reagent replacement interval and conditioning, with point-by-point checks on: Ex certification and GMP traceability of records.
Operating Outcome
In service the system performs as intended — steady reagent and consumable supply, making upkeep predictable, with routine maintenance falling noticeably.
Power & Energy · HACH-ONLINE — Application Case
Application Context
boiler feedwater, turbine condensate and FGD/DeNOx systems is a typical Power & Energy duty. The project started from HACH-ONLINE, focusing on continuous outfall monitoring with Hach online instruments.
Process Constraints
drift under high temperature and pressure, with very costly unplanned outages, and high temperature and pressure with both steam and saline water only amplify it. The HACH-ONLINE specification must therefore cover configured by monitoring range, reagent replacement interval and conditioning.
Delivery Approach
Delivery was handled as the online water-analysis line — integrating around each online monitoring point. Execution centred on configured by monitoring range, reagent replacement interval and conditioning, together with compliance with power-industry codes and unit availability targets.
Field Results
Operation shows the benefit clearly — steady reagent and consumable supply, making upkeep predictable, with confidence in this measuring point improving markedly.
Related Products
HACHSigma total phosphorus / orthophosphate analyzer
sc dichromate COD online analyzer
HACHOnline water hardness analyzer
HACHMulti-parameter automatic water analyzer
HACHMulti-parameter online water analyzer
HACHOnline silica analyzer
HACHAmmonia nitrogen analyzer
HACH